water flow meter-metal tube variable area flowmeter features high precision, low drift, strong capability of anti-interference

water flow meter-MTF metal tube variable area flowmeter

MTF intelligent metal tube float flowrate meter is developed and manufactured one of intelligent series instrument. It is one kind of a combined that integrates analog circuit, digital circuit and microprocessor. It converts fluid flow signal into corresponding analog voltage signal which then is changed into 4-20mA two wire system current output signal on which HART protocol communication is superimposed. The meter adopts HART bus intelligent transmitter. Because of HART bus technology, it features high precision, low drift, strong capability of anti-interference, etc. It can realize functions for remote configuration, monitoring, maintenance and calibration, etc. It can compose production procedural measurement, monitoring control system. The meter can be classified into basic type ( horizontal mounting type, vertical mounting type), jacket type, damping type. It shapes a beautiful appearance that shows the modern fashionable case design with operating chambers separated.

Main performance and technical parametersPerformance

Setting various variables: To set the meter with the local switch, hand-hold communicator and debug software (program).

LCD indication: Indicate relative variables

Read procedural variables: Two-wire system 4-20mA current output on which HART protocol communication (signal).

Read and diagnose information: To provide alarm for exceeding flowrate range and diagnose different failures. Featuring anti-shock and magnetic interference resistance functions, etc.

Having functions for zero and flowrate range migration.

Protection class: IP65.

Flange standard: JB/T82.1-94, JB/T82.2-94(raised face flange).

Qualifications

Operating voltage: DC24V.

Power supply inlet: M20X1.5.

Input impedance: more than 100MΩ.

Standard load resistance: 250Ω.

Nominal pressure: 1.6, 2.5, 4.0, 6.4MPa.

Precision: 1.5%FS, (for diameter DN15, precision is 2.0%).

Operating temperature: -35~200°C.

Ambient temperature: -35~70°C.

Damping period: (0-199s).

Explosion-proof type, explosion-proof mark, explosion-proof certificate

Explosion-proof type

Explosion-proof mark

explosion-proof certificate

Intrinsically safe 

ExibIICT4

GYB02163

Isolating explosion

ExdIIBT4

GYB02162

Measuring range

for water, 2.5--100000L/h (20°C),  for air, 0.07--1800m3/h(0.1013MPa, 20°C).

Applicable fluid viscosity

DN15:η less than 5mPa.s(FZ15.1,FZ15.2,FZ15.3)

η less than 30mpa.s(FZ15.4,FZ15.5,FZ15.6,FZ15.7,FZ15.8) 

DN25: η less than 250mPa.s  

DN50-100: η less than 300mPa.s 